Lab Care Furniture
Atladra, Vadodara, Gujarat
GST No. 24BGFPS4355E1ZF
Send E-mail
Call 0804876222780% Response Rate
Home  » Our Products  » 

Fume Hood


Pioneers in the industry, we offer bench type laboratory fume hood, distillation fume hood, chemical fume hood and walk in fume hood from India.

  • +7
Get Best Quote

Approx. Rs 1.62 Lakh / Unit Get Latest Price

Product Details:
Usage/ApplicationLaboratory
Frequency50/60 Hz
External Size1800 x 940 x 2150 mm
Internal Size1600 x 670 x 1250 mm
Voltage Supply110 / 220 V +- 10%
Product TypeBench Type

Fume hood super structure should consist of

-         Frame structure in atleast 1.6 mm thick galvanized iron construction.

-        Inner lining, rear baffles and top baffles in 6mm thick fire retardant property lab grade Phenolic resin construction of any reputed make.

-        Outer panels of the super structure should be in atleast 1.2mm thick high finished GI construction, duly coated with epoxy powder.

-        Front post should be broad enough to mount service fittings and should be in fabricated out of atleast

1.2 mm thick high finished GI construction, duly coated with epoxy powder.

-        Air Foil should be in Flip-on type design and should be fabricated out of 1.6 mm thick SS 316 construction. Air foil should be with hard teflon coating for extra ordinary chemical resistant.

-        Exhaust outlet should be provided with streamlined shaped duct collar in PP construction.

-        The front sash should be in Combination type design, 5 mm thick toughened glass construction with full length sash handle in anodized Al construction duly coated with epoxy powder.

-        Sash movement mechanism should be provided with multi core SS wire having dia. not less than 3.0 mm. The wire pulleys should be provided with ball bearings in SS construction for smooth movement of the sash.

-        Super structure should be provided with LED lights of sufficient capacity for proper illumination within the hood chamber. The light intensity at work top level should be at least 300 lux.

 

Fume hood work top in granite construction having thickness not less than 18 mm. The work top shall be provided with raised rails on all 4 sides to prevent spillage and should also be provided with a cup sink in PP construction on rear side corner.

 

All the service valves should be in forged brass construction with spindle in SS construction. Valves should be provide with angular shaped serrated nozzles also in brass construction with epoxy powder coating. The valve knobs should be with color sign as per DIN standard. All the valves should be provided with service pipe connection at least 1.5-meter-long in following construction.

 


Yes! I am Interested


REQUEST CALLBACK
Get Best Quote

Approx. Rs 1.9 Lakh / Unit Get Latest Price

Product Details:
MaterialMild Steel, Stainless Steel
Air Velocity0.3 - 0.8 m/s
UsageLaboratory
Gross Weight230 kg - 420 kg
Height4, 5, 6, 7 Feet and Customized

Laboratory metallic furniture must be tested as per SEFA standards in SEFA Approved labs with latest

2016 Guidelines published by SEFA.

 


Yes! I am Interested


REQUEST CALLBACK
  • +4
Get Best Quote

Approx. Rs 1.45 Lakh / Unit Get Latest Price

Product Details:
MaterialMild Steel
Size1200/1500/1800 x 940x 2250 mm
UsageChemical
Air Volume1750 m3/h

1.     Fume Hood Acid Storage Cabinets:

Acid storage fume hood cabinets shall utilize the same thickness of steel and construction features as other base cabinets except they shall be completely lined with a Polyethylene corrosion resistant liner. The liner shall be 5mm thick, Each door shall have a set of louvers at the top and bottom, and have a 3mm sheet polyethylene liner. Where specified, each cabinet shall be vented into the fume hood with a 40mm vent pipe. It should provide a positive airflow directly into the fume hood exhaust system. Where specified or shown on drawings, epoxy coated wire shelf supported by integral brackets shall be built into the Polyethylene liner.

 


Yes! I am Interested


REQUEST CALLBACK
Get Best Quote

Approx. Rs 1.7 Lakh / Unit Get Latest Price

Product Details:
Air Velocity0.3 - 0.8 m/s
UsageChemical, Laboratory
Dimension1200 x 800 x 2350 mm Approx
Temperature Range-30 to 50 Degree C

Fume hood super structure should consist of

-         Frame structure in atleast 1.6 mm thick galvanized iron construction.

-        Inner lining, rear baffles and top baffles in 6mm thick fire retardant property lab grade Phenolic resin construction of any reputed make.

-        Outer panels of the super structure should be in atleast 1.2mm thick high finished GI construction, duly coated with epoxy powder.

-        Front post should be broad enough to mount service fittings and should be in fabricated out of atleast

1.2 mm thick high finished GI construction, duly coated with epoxy powder.

-        Air Foil should be in Flip-on type design and should be fabricated out of 1.6 mm thick SS 316 construction. Air foil should be with hard teflon coating for extra ordinary chemical resistant.

-        Exhaust outlet should be provided with streamlined shaped duct collar in PP construction.

-        The front sash should be in Combination type design, 5 mm thick toughened glass construction with full length sash handle in anodized Al construction duly coated with epoxy powder.

-        Sash movement mechanism should be provided with multi core SS wire having dia. not less than 3.0 mm. The wire pulleys should be provided with ball bearings in SS construction for smooth movement of the sash.

-        Super structure should be provided with LED lights of sufficient capacity for proper illumination within the hood chamber. The light intensity at work top level should be at least 300 lux.

 


Yes! I am Interested


 